K41091 Micro Carbon Residue and Ash Tester
Operation and Instruction Manual
K41091-Manual
-
12-
Figure 11: Cleaning Cable_Top View
Grab the end of the cleaning cable from
where the condensate tank normally sits.
See Figure 12 below:
Figure 12: Cleaning Cable_Bottom View
With a back and forth movement and
directing the cable in all direction
simultaneously, clean the pipe of all
calamine deposits.
Check the state of the chimney and chimney
extension pipe for calamine deposits and
clean in the same manner.
6.2.3 Temperature Check of the Furnace
To check the temperature of the furnace it is
necessary for a control thermocouple to be in
place. In order to make this possible, the hot lid
has a small thermocouple port located along the
outside. If the actual temperature of the furnace
does not correlate to the reading on the
instrument, check the calibration of the furnace
thermocouple.
See
the
location
of
the
thermocouple port in Figure 13 below:
Figure 13: Hot Lid Thermocouple Port