K41091 Micro Carbon Residue and Ash Tester
Operation and Instruction Manual
K41091-Manual
-
9-
The maximum time duration is 240 minutes per
segment. The type of gas flow to choose from is
listed in section 5.1 of this manual.
NOTE:
The Temperature Rate must not exceed
15°C per minute in order to maintain the condition
of the Furnace.
Creation or Modification:
1.
From the main menu press the
“Go”
key
located at the bottom left corner of the screen.
2.
Next press the
“Program”
key at the bottom
right corner of the screen.
3.
Enter the number of the program to modify or
create. This will bring you to the screen
illustrated in Figure 9 below.
Figure 9: Program Parameter Screen
4.
Pressing the
“Validate the Parameter”
key
displays the type of gas and flow available.
5. Alphanumeric Keyboard Function.
Pressing
the
“A…Z”
key displays a table with column
headings and lines allowing the user to select
a letter, figure or symbol for inputting a sample
ID or naming a newly created program. Please
see Figure 10 below:
6. Weight Input Function.
This function makes
it possible to enter the weights of the samples
so that the apparatus can make a calculation
after the weights entered at the end of the
test. When entering a weight value, the
numbers will start on the operators left and
scroll to the right as additional numbers are
entered. To enter the desired value it may be
required to enter a number of “Zeros” after the
value has been entered. The standard number
of significant figures is shown below:
XX.XXXXX G
For example, to enter 2.45 G, the user must
enter 2, then 4, then 5 followed by three
Zeros. This places the two (2) in the single
digit place resulting in the desired value.
7. Set Point Function.
This function allows the
user to input the idle set point temperature of
the furnace between tests. This value is
limited to the range from 20 to 80°C
8. Standby Function.
This function allows the
user to shut down the apparatus via the
control panel without having to switch off the
circuit breaker or push the emergency stop
button. This method of shut down avoids
restarting the circuit breaker at the time of the
next power up. Upon activation of this
function, the instrument automatically shuts
down after a few seconds.
5.3.2 Software Configuration Functions
The following functions are accessible via the
Software Function
.
From the main menu press the
“Configuration
Key
”
located at the bottom right corner. During
this functions
’ first use a password is not required
to enter this menu. A password must be created in
the “New Password” function screen to implement
this security feature. Once in the Software
Function Menu, press next and enter the “New
Password Function” screen. See step 5 below for
further detail on this function.
1. Buzzer Function.
This function allows the
user to choose between a continuous buzzer
that sounds at the end of the test until the
operator turns it off or a buzzer that sounds for
3 minutes.
2. Stop Ventilation Function.
This function is
automatically set to 20°C above the Idle Set
Point Temperature value. The minimum value
for this set point is 80°C. The fan starts at
500°C and stops at the Idle set point plus
20°C.
3. Stop Safety Function.
This function allows
the user to set a safety temperature limit.
This
temperature is measured on the Furnace
temperature probe. If a malfunction occurs
and the temperature of the furnace exceeds
this value, the microprocessor cuts the heating
Summary of Contents for K41091