Koehler K13902 Operation And Instruction Manual Download Page 10

K13992 Cleveland Open Cup Flash Point Tester 

Operation and Instruction Manual 

 

 

 

K13992-Manual

    

 

-

6-

 

 

1  Introduction  

 

The Koehler K13992 Cleveland Open Cup Flash 

Point  Tester  is  the  latest  design  for  accurately 
determining  flash  and  fire  point  temperatures  of 
viscous  petroleum  products  over  an  extended 
temperature  range  according  to  the  ASTM  D92 
test method and related test specifications.  

 
This  manual  provides  important  information 

regarding  safety,  technical  reference,  installation 
requirements,  operating  condition  specifications, 
user  facility  resource  requirements,  and  operating 
instructions  for  the  Cleveland  Open  Cup  Flash 
Point Tester.  This manual should also be used in 
conjunction  with  applicable  published  laboratory 
procedures.  Information  on  these  procedures  is 
given in section 1.2. 
 

1.1  Koehler's 

Commitment 

to 

Our 

Customers 

 

Providing  quality  testing  instrumentation  and 

technical support services for research and testing 
laboratories  has  been  our  specialty  for  more  than 
50  years.  At  Koehler,  the  primary  focus  of  our 
business  is  providing  you  with  the  full  support  of 
your  laboratory  testing  needs.  Our  products  are 
backed  by  our  staff  of  technically  knowledgeable, 
trained  specialists  who  are  experienced  in  both 
petroleum products testing and instrument service 
to  better  understand  your  requirements  and 
provide  you  with  the  best  solutions.  You  can 
depend on Koehler for a full range of accurate and 
reliable instrumentation as well as support for your 
laboratory  testing  programs.  Please  do  not 
hesitate  to  contact  us  at  any  time  with  your 
inquiries  about  equipment,  tests,  or  technical 
support. 

 

Toll Free: 1-800-878-9070 (US only) 
Tel: +1 631 589 3800 
Fax: +1 631 589 3815 
 
Email: [email protected] 
http://www.koehlerinstrument.com 

 
 
 
 
 
 
 
 

1.2  Recommended 

Resources 

and 

Publications 

 

1. 

American  Society  for  Testing  and  Materials 
(ASTM) 
100 Barr Harbor Drive 
West  Conshohocken,  Pennsylvania  19428-
2959, USA 
Tel: +1 610 832 9500  
Fax: +1 610 832 9555 

http://www.astm.org

 

email: [email protected] 

 

ASTM Publication: 

  ASTM  D92:  Standard  Test  Method  for  Flash 

and  Fire  Points  by  Cleveland  Open  Cup 
Tester 

  ASTM 

D6074: 

Standard 

Guide 

for 

Characterizing  Hydrocarbon  Lubricant  Base 
Oils 

  ASTM  D6158:  Standard  Specification  for 

Mineral Hydraulic Oils 

 

2. 

International Organization for Standardization 
(ISO) 
1, rue de Varembé 
Case postale 56 
CH-1211 Geneva 20, Switzerland 
Tel: 41 22 749 01 11  
Fax: 41 22 733 34 30 
http://www.iso.org 

 

ISO Publication: 

  ISO  2592:  Determination  of  Flash  and  Fire 

Points- Cleveland Open Cup Method 

 

3. 

American  Association  of  State  Highway  and 
Transport Officials (AASHTO) 
444 North Capital Street N.W., Suite 249 
Washington, D.C. 20001 
Tel: (202) 624-5800 
Fax: (202) 624-5806 
E-mail[email protected] 
 

AASHTO Publication: 

  AASHTO  T48:  Flash  and  Fire  Points  by 

Cleveland Open Cup 

 

4. 

Energy Institute (IP) 
61 New Cavendish Street 
London, WIM 8AR, United Kingdom 
Tel: 44 (0)20 7467 7100  
Fax: 44 (0)20 7255 1472 
http://www.energyinstpubs.org.uk/ 

 

Summary of Contents for K13902

Page 1: ... Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing K13902 K13992 Cleveland Open Cup Flash Point Tester ...

Page 2: ......

Page 3: ...pplicable standards set forth in this certification Specifications ASTM D92 ASTM D6074 AASHTO T48 ASN Z 11 6 IP 36 ISO 2592 DIN 51376 FTM 791 1103 FTM 141 4294 This unit is tested before it leaves the factory to ensure total functionality and compliance to the above specifications and ASTM standards Test and inspection records are on file for verification Jesse Kelly Application Engineer Koehler I...

Page 4: ......

Page 5: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 1 ...

Page 6: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 2 ...

Page 7: ...c information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Reporting Koehler will not sell directly to end users in the EU and so has no responsibility to register within each EU state and to make annual reports Koehler declares that this responsibility is born...

Page 8: ...rately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is greater than 10 square centimeters about 4 sq in Toner cartridges liquid and pasty as well as color toner Chlorofluorocarbons CFC hydrochlorofluorocarbons HCFC or hydrofluorocarbons HFC hydrocarbons ...

Page 9: ... 2 Additional Accessories Required for Testing 8 3 3 Unpacking 8 4 Setup 9 4 1 Instrument Descriptions 9 5 Operation 11 5 1 Test Procedure 11 6 Safety Features 12 6 1 Over Power Protection 12 7 Maintenance 12 7 1 Routine Maintenance 12 7 2 Instrument Cleaning 12 7 3 Replacement Parts 12 8 Wiring Diagrams 13 8 1 220 240V Unit Wiring 13 9 Troubleshooting 14 9 1 Unit does not power up 14 9 2 Unit is ...

Page 10: ... You can depend on Koehler for a full range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or technical support Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 Email info koehlerinstrument com http www koehlerinstrument com 1 2 Recommen...

Page 11: ...ed with the use of this equipment It is the responsibility of any user of this equipment to investigate research and establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use Equipment Modifications and Replacement Parts Any modification or alteration of this equipment from that of factory specifications is NOT recommended because it v...

Page 12: ...g for correct orientation of instrument e g This Side Up 3 Carefully open top of box with box cutter and remove packing foam insert 4 Extract instrument and place on suitable cart for transportation to work area lab bench WARNING Be sure two or more individuals are available for extracting and lifting instrument from box to cart and from cart to bench Individuals must lift in accordance to proper ...

Page 13: ... or blow out the test flame as this will affect results of the test Gas Supply Connect the gas inlet to a regulated low pressure gas supply 0 5 to 1 0 psi Use only propane LPG or natural gas Do not use direct unregulated pressure from an LPG tank NOTE Be sure Flame Control Switch Figure 1 Item 3 is in the OFF position when Test Flame Applicator is not in use Power Connect the line cords to properl...

Page 14: ...per test method specifications 6 Thermometer Holder This holder is used to place the thermometer in the test cup to measure the temperature of the test specimen The holder can be tilted back to a position away from the test cup when instrument is not performing a test 7 Cleveland Open Flash Cup The brass test cup contains the test specimen and is built with a heat resistant handle An indicator lin...

Page 15: ...ases 5 to 17 C 9 to 30 F min 10 When the test specimen temperature is approximately 56 C 100 F below the expected flash point decrease the rate so that the temperature rise during the last 28 C 50 F before the flash point is 5 to 6 C 9 to 11 F min 11 Apply the test flame when the temperature of the test specimen is approximately 28 C below the expected flash point and each time thereafter at a tem...

Page 16: ... over the course of time some instrument parts may need to be replaced When ordering replacement part s please provide the model number serial number and product shipment date of your equipment so that we can ensure you will receive the proper replacement part s 7 2 Instrument Cleaning To clean the instrument s exterior which includes all painted surfaces either a solution of soap and water or lab...

Page 17: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 13 8 Wiring Diagrams 8 1 220 240V Unit Wiring ...

Page 18: ...e consistent with an indoor laboratory environment This testing equipment should not be subjected to extremes of temperature and or moisture This equipment was shipped from the factory in a corrugated cardboard container If long term storage is anticipated re packing the instrument in a water resistant container is recommended to ensure equipment safety and longevity 12 Warranty We at Koehler woul...

Page 19: ... when you return the material to our plant Shipping containers without an RA number prominently displayed will be returned to the sender Goods must be returned freight prepaid Returns will be subject to a restocking charge the application of which will depend upon the circumstances necessitating the return Some returns cannot be authorized including certain products purchased from outside vendors ...

Page 20: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 16 Notes ...

Page 21: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 17 Notes ...

Page 22: ...K13992 Cleveland Open Cup Flash Point Tester Operation and Instruction Manual K13992 Manual 18 Notes ...

Reviews: