9-26
1F1976
August 1, 2003
Key Operator Functions
3.
Select the
Procedure Code
or
Procedure
in the list that you want to
edit.
4.
Select
Update Procedure.
Using a HIS/RIS System
To configure Procedure Mapping to work with your HIS/RIS System, make
sure that the code you enter for the Procedure matches the code that comes
from the RIS. You can use any number for the Procedure Code; however, it
will only work automatically with the HIS/RIS when the codes match exactly.
For example, if you had a HIS/RIS procedure “2-view Chest” code 102 and
you want to create a procedure on the CR System called 2-view Chest:
1.
At the
Key Operator Functions
menu, select
System Configuration
then select
Workflow Optimization.
2.
Select
Procedure Mapping
.
3.
Create a procedure named 2-view Chest and enter a code of 102.
4.
Select the icons you want to be displayed when a 2-view chest procedure
is ordered, such as an AP and LAT chest.
When the worklist on the CR System is updated and the RIS code 102 is
received, the AP chest and LAT chest image icons are automatically
displayed.
Deleting a Procedure
1.
At the
Key Operator Functions
menu, select
System Configuration
then select
Workflow Optimization.
2.
Select
Procedure Mapping.
3.
Select the procedure that you want to delete and select
Delete
Procedure
.
4.
Select
Yes
on pop-up screen.
Procedure Mapping
Using the Remote Key
Operator
You can use the Remote Key Operator to create or change Procedure
Mapping with these features:
•
Add—
Enter a new Procedure Code (required) and Procedure Name and
use the drop-down menus to map the associated exams.
•
Find—
Search for a Procedure Code or Procedure Name.
•
Update—
Edit the Procedure Code or Procedure Name.
•
Copy From—
Copy from an existing Procedure Code to a new or existing
Procedure Mapping entry.
•
Delete—
Delete the procedure mapping for any existing procedure.
Summary of Contents for DirectView CR 500 System
Page 1: ...Kodak DirectView CR 500 System User s Guide H195 0016AC...
Page 10: ...viii August 1 2003 Table of Contents...
Page 14: ...1 4 1F1976 August 1 2003 Safety and Related Information Labels H195_0001DC...
Page 15: ...Safety and Related Information August 1 2003 1F1976 1 5 H195_0002DC...
Page 16: ...1 6 1F1976 August 1 2003 Safety and Related Information H195_0003DC...
Page 28: ...2 6 1F1976 August 1 2003 Overview...
Page 36: ...3 8 1F1976 August 1 2003 Operation and Workflow...
Page 46: ...4 10 1F1976 August 1 2003 Entering Exam Data...
Page 52: ...5 6 1F1976 August 1 2003 Scanning Viewing and Managing Images Reject Images...
Page 80: ...5 34 1F1976 August 1 2003 Scanning Viewing and Managing Images...
Page 92: ...6 12 1F1976 August 1 2003 Maintaining Image Quality...
Page 97: ...Troubleshooting August 1 2003 1F1976 7 5 Clear Pending Images...
Page 108: ...8 10 1F1976 August 1 2003 Maintaining Equipment and Cassettes...
Page 189: ...Key Operator Functions August 1 2003 1F1976 9 81 Number 2 Number 3 Number 4 Number 5...
Page 190: ...9 82 1F1976 August 1 2003 Key Operator Functions Number 6 Number 7 Number 8 Number 9...
Page 191: ...Key Operator Functions August 1 2003 1F1976 9 83 Number A Number B Number C Number D...
Page 194: ...9 86 1F1976 August 1 2003 Key Operator Functions Clear All Suffix Clear All Prefix Test...
Page 202: ...9 94 1F1976 August 1 2003 Key Operator Functions...
Page 218: ...10 16 1F1976 August 1 2003 Remote Operations Panel...
Page 226: ...B 4 1F1976 August 1 2003 Printing Exceptions...
Page 235: ......