8-6
1F1976
August 1, 2003
Maintaining Equipment and Cassettes
Replacing the Flexible Phosphor Screen
After cleaning the phosphor screen:
1.
Position the cassette with the Tube Side up. See the
.
2.
Place the tip of a tool, such as a small screwdriver, or ballpoint pen in
the oval hole on the side of the cassette and push in to release the latch.
Move the latch forward to open the cassette.
3.
Place the white (phosphor) side of the screen up so that it faces the
cassette’s Tube Side panel.
NOTE: Use only KODAK Flexible Phosphor Screens for KODAK DIRECTVIEW
CR 500 Systems, available separately. Match the screen resolution to
the resolution indicated on the cassette label: (GP) General Purpose
or (HR) High Resolution
4.
Carefully insert the screen all the way into the cassette.
NOTE: For proper operation and calibration of the 35 x 35 cm cassette,
always insert the screen into the cassette so that the black writing is at
the cassette door or back edge. It is possible to insert this screen into
the cassette in any of four orientations, and if you fail to insert the
screen correctly, it could result in both calibration and functional
failures:
– The scan calibration procedure may fail causing an error
message.
– The screen may not feed properly from the cassette, causing an
error message and preventing any exposed image on the screen
from being processed.
5.
Place the tool into the round hole and pull the latch back as far as
possible to close the cassette.
Cassette Cautions
CAUTION
:
You cannot ship cassettes contaminated with blood or other
body fluids to an Eastman Kodak Company facility for
evaluation unless the cassettes have been decontaminated.
Please utilize “universal precautions” and decontaminate the
cassettes with either an EPA registered tuberculocidal (list B)
or diluted bleach (1:10 with water) solution prior to shipping.
Summary of Contents for DirectView CR 500 System
Page 1: ...Kodak DirectView CR 500 System User s Guide H195 0016AC...
Page 10: ...viii August 1 2003 Table of Contents...
Page 14: ...1 4 1F1976 August 1 2003 Safety and Related Information Labels H195_0001DC...
Page 15: ...Safety and Related Information August 1 2003 1F1976 1 5 H195_0002DC...
Page 16: ...1 6 1F1976 August 1 2003 Safety and Related Information H195_0003DC...
Page 28: ...2 6 1F1976 August 1 2003 Overview...
Page 36: ...3 8 1F1976 August 1 2003 Operation and Workflow...
Page 46: ...4 10 1F1976 August 1 2003 Entering Exam Data...
Page 52: ...5 6 1F1976 August 1 2003 Scanning Viewing and Managing Images Reject Images...
Page 80: ...5 34 1F1976 August 1 2003 Scanning Viewing and Managing Images...
Page 92: ...6 12 1F1976 August 1 2003 Maintaining Image Quality...
Page 97: ...Troubleshooting August 1 2003 1F1976 7 5 Clear Pending Images...
Page 108: ...8 10 1F1976 August 1 2003 Maintaining Equipment and Cassettes...
Page 189: ...Key Operator Functions August 1 2003 1F1976 9 81 Number 2 Number 3 Number 4 Number 5...
Page 190: ...9 82 1F1976 August 1 2003 Key Operator Functions Number 6 Number 7 Number 8 Number 9...
Page 191: ...Key Operator Functions August 1 2003 1F1976 9 83 Number A Number B Number C Number D...
Page 194: ...9 86 1F1976 August 1 2003 Key Operator Functions Clear All Suffix Clear All Prefix Test...
Page 202: ...9 94 1F1976 August 1 2003 Key Operator Functions...
Page 218: ...10 16 1F1976 August 1 2003 Remote Operations Panel...
Page 226: ...B 4 1F1976 August 1 2003 Printing Exceptions...
Page 235: ......