background image

Part No. 8G8528

15 April 09

KODAK Medical X-ray Processor 2000

Model 212 / 220-240V, 50 Hz

Model 213 / 220-240V, 60 Hz

Service Manual

Summary of Contents for 212

Page 1: ...Part No 8G8528 15 April 09 KODAK Medical X ray Processor 2000 Model 212 220 240V 50 Hz Model 213 220 240V 60 Hz Service Manual ...

Page 2: ... bath temperature is not reached 8 The developer bath temperature is higher than the set temperature the display shows two crossbars the developer temperature button blinks 8 The developer bath temperature too high or too low display shows values of 20 C or 42 C 8 The calibration of the bath temperature actual bath temperature is different from the displayed value 8 Calibration process 9 Film defe...

Page 3: ...ii Water 10 Rinsing water does not flow ventilation runs 10 Water tank overflows 10 Spare Parts 11 Wires and cables 12 Standard roller racks 15 Electrical Diagrams 18 ...

Page 4: ... information Carestream Health Inc reserves the right to change this information without notice and makes no warranty express or implied with respect to this information Carestream Health shall not be liable for any loss or damage including consequential or special damages resulting from any use of this information even if loss or damage is caused by Carestream Health s negligence or other fault ...

Page 5: ... Support Type Number Hours Phone 49 7062 9255 0 Normal Business Hours Central European Time Berlin Paris Rome Monday Thursday 8 am to 5 pm Friday 8 to 12 am noon Direct Mail PROTEC GmbH Co KG Lichtenberger Strasse 35 D 71720 Oberstenfeld Germany E Mail service protec med com ...

Page 6: ...3 Dimensions Dimensions Depending on the model type the MXP 2000 has either an open stand or a closed cabinet Transport To carry hold here To carry hold ...

Page 7: ... Installation Data Installation Data Through the wall mounting film infeed Machine in light room Through the wall mounting set 1280 0 0000 Wall breakthrough for film input Dimensions in cm Working space ...

Page 8: ...n Data Through the wall mounting film output Film output to the light room Through the wall mounting set 1281 0 0000 Wall breakthrough for film exit Dimensions in cm Working space Door width min 69 Film exit height ...

Page 9: ...d externally 10 Wall breakthrough for film exit through the wall infeed page 4 The processor is installed outside the darkroom the film exits into a light room For the through the wall mounting infeed alteration assembly kit no 1280 0 0000 is necessary 11 Wall breakthrough for film exit through the wall output page 5 The processor is installed inside the darkroom the film exits through the wall to...

Page 10: ...splay again Disconnect the temperature sensor dryer X24 and check the display again 3 Drive 3 1 Machine does not start automatically When the machine is switched on place a film in the infeed until it reaches the pull in rollers If necessary move the film sideways to activate the light barrier If the display shows two bars with decimal points then the light barrier is in order Check each of the fo...

Page 11: ...t reached 4 3 The developer bath temperature is higher than the set temperature the display shows two crossbars the developer temperature button blinks See above Check the current on power PCB at clip X13 There may be no current If current is registered replace the electronics 4 4 The developer bath temperature too high or too low display shows values of 20 C or 42 C Check the temperature sensor T...

Page 12: ... are damaged replace them Pressure marks are caused by contaminated or damaged rollers If necessary replace the rollers 6 Replenishment 6 1 The replenishment pump does not work or works insufficiently 7 Dryer 7 1 The display shows error message E4 The following parts should be checked and if faulty be replaced Power PCB in standby no voltage at X12 Temperature sensor see The dryer temperature disp...

Page 13: ...Measure the voltage at X24 on the power PCB between pin 3 green and pin 2 brown The value should read 0 32 V at a temperature of 32 C Enter a dryer temperature of 35 C Check the voltage of X12 on the power PCB If no voltage can be registered replace the control unit 7 6 Dryer temperature cannot be reached or is too high display shows values approximately 30 C and 75 C Check the temperature sensor ...

Page 14: ...r inlet tube 11 1196 0 0010 Processor base stand open 12 1190 0 1201 Control panel foil 2018 0 0005 Tube 4x1mm green 2018 0 0021 Tube 9x2mm red clear 2018 0 0022 Tube 9x2mm blue clear 2018 0 0007 Tube 16x4mm clear reinforced 2018 0 0008 Tube 10x2mm red reinforced 2018 0 0009 Tube 10x2mm blue reinforced 2018 0 0012 Tube 10x2mm clear reinforced 2022 0 0014 Tube clamp Snap 2022 0 0019 Wire tube clamp...

Page 15: ...rmer power PCB 20V 1190 0 0112 Power PCB transformer 230V UL 1190 0 0110 Power PCB controller PCB 1190 0 0111 Motor controller PCB 1190 0 0114 Fuse holder controller PCB UL 2004 0 0010 Electrical power lead 2004 0 0016 Electrical power lead UL 2004 0 0012 Main switch fuse holder UL ...

Page 16: ...motor new 31 1190 0 1102 Motor bracket variation 2 32 1190 0 1503 Drive shaft variation 2 33 1170 0 1502 Bearing block 34 1170 0 1503 Worm gear 36 0170 0 2400 Micro switch Cover 0170 4 2400 Micro switch Cover UL 37 2007 0 0010 Operator for micro switch 38 1190 0 0902 Bracket for cover switch variation 2 40 0196 0 0900 Controller PCB 41 2028 0 0023 Power switch 42 2010 0 0004 Fuse holder 2010 0 001...

Page 17: ...sor same for bath and dryer heating 54 1190 0 1302 Plate for air channel 55 2003 5 0002 Heating element bath heating 2003 7 0002 Heating element 800 W UL 56 1190 0 2101 Heat exchanger 57 2005 0 0005 Temperature limiter 59 1190 0 4100 Condensation cover E 60 1190 0 0103 Light barrier support 61 2030 0 0014 Rubber inlet 62 63 0190 0 0800 Light barrier holder compl 64 1190 0 0107 Light barrier casing...

Page 18: ...15 Spare Parts Standard roller racks ...

Page 19: ...ler ground 75 1101 0 0307 Rubber roller 77a 1190 0 0302 Flat spring 68 77b 1170 0 0304 Flat spring 55 77c 1170 0 0303 Flat spring 36 77d 1101 0 0315 Flat spring 30 77e 1190 0 0303 Flat spring 26 79 1190 0 0310 Drive shaft rack C 2 80 1101 0 0316 Shaft 81 1101 0 0311 Distance bar 82 1101 0 0303 Gear t 32 D hole 83 1101 0 0304 Gear t 32 R hole 84 1170 0 0302 Worm wheel 85 1101 0 0302 Gear t 16 D hol...

Page 20: ...02 1253 0 0001 Catch basket 103 0281 0 0202 Mounting kit 0281 0 0204 Strap 104 0281 0 0203 Blind plate 2 Position Order No Description 110 115 1280 0 0000 Mounting set Through the wall mounting set film feed 110 0280 0 0102 Light protection 111 0280 0 0103 Light protection left 112 0280 0 0104 Light protection right 113 0280 0 0105 Wall plate small 114 0280 0 0202 Mounting kit ...

Reviews: