Page 82 of 105
OPERATING MANUAL HPC & UMC4
14.11.1 Simulation on/off
The
Simulation on/off
function allows the operator to activate or deactivate simulation. If simulation is
activated, all output signals will be generated based on the selected type of simulation. The peripherals
connected to the device can be tested without a flowing product. Press
„Enter“ to display the current sta-
tus.
Simulation
[off]
As mentioned in Section 13.4.3.1 Selection window/make a selection, the operator toggle between the
“on” and “off.”
Simulation will be deactivated automatically if the operator switched the device off or did not touch any
control unit keys for about 10 minutes.
14.11.2 Direct simulation
This function allows the operator to define whether simulation is comprised of the measurement of the
three physical values mass flow, density and temperature or whether the outputs will be set directly.
Press
„Enter“ to display the selected type of simulation.
Simulation
[direct]
As mentioned in Section 13.4.3.1 Selection window/make a selection, the operator can choose between
the following settings:
Direct
pulse and current outputs are programmed directly
QM, D, T
a measurement is simulated
If “direct” simulation is activated, any output will perform based on the settings described in Sections
14.11.4.1 Status output simulation to 14.11.4.4 Simulation current output I2. It is therefore recommended
that the settings be defined before starting simulation. They can then be purposefully changed during
simulation.
The status of the outputs during measured value simulation based on the setting “QM, D, T” depends on
the selected simulation values of these three variables, the measuring range settings and the assignment
of the outputs. If, for example, the pulse output is assigned to volume measurement, it will be affected by
all three simulation values at the same time [V
≈ QM (T) / D (T)].
Simulation will be deactivated automatically if the operator switched the device off or did not touch
any control unit keys for about 10 minutes.