Kobold DUC-ZUWD Quick Start Download Page 1

DUC-ZUWD      Quickstart   

 

 

 

 

 

Prepare your wall thickness gauge /Keypad definition 

 
The delivery is usually ready to use; batteries (AA). 

 
Switch on / off 

Backlight on / off 

Scroll down /

 

 

 

 

        

decrease 

Exit 

           

Menu / Select 

 
Enter              

Calibration    

 

Scroll up/ 

 

 

 

 

  

Increase 

 

3 steps to parameterize your wall thickness gauge 

 

1.  Please check if the sensor probe type is set correctly (standard type ‘N07’) 
2.  Make calibration using integrated calibration sample 
3.  Set value for speed of sound of your pipe material. 
4.  Place probe on the pipe and get reading for wall thickness 

 

Please make sure to put some coupling grease on the sensor tip before measurement. 
Calibration should be repeated each time you start new measurement. 
Calibration should be done at speed of sound of 5920 m/s 

 

 
 
 
Put some coupling grease (Magnalube) on the probe tip (Pic.1). 
Put sensor on the integrated calibration sample (Pic.2) which you find in 
the lowermost right corner of the instrument. 

Press button 

. Display will show 4.00 mm. Remove the probe.  

Your DUC-ZUWD is now calibrated. Now set the speed of sound of media 

of pipe material. Press button 

 (4x) to go to menu for setting of speed 

of sound. (lowermost line, right corner), see pic.3. 

The value can be changed by using buttons 

 

Values for typical  

materials can be obtained from the adjacent table. Put sensor tip on your pipe  
and take readings from display, see pic.4. 

 
 
 
 
 
 
 
 
 
 
 

Connect 
the signal 
cables of 
the 
transducer 
to the 
thickness 
gauge 

Calibration sample 

Pic. 1 

Pic. 2 

Pic. 3 

Pic. 4 