61
Parameter Setting: The Start-Up Program
At the end of the parameter-setting procedure, a ”Start-up” line appears in the
”Installation” menu.
When you are sure to have set all parameters, select ”Yes” to confirm.
Now the pumps perform the number of stroke movements required for filling
the media tubes completely.
The necessary rinsing cycles are automatically started.
The buffer pumps require approx. 1 stroke to fill the pump and approx.
9 strokes to fill the tubing.
13 Commissioning
Start-Up Program
NOTICE
When the media connection is longer than 10 m, three further pump strokes
are required to fill the tubes.
Sensor Dismount Guard
To check the sensor dismount guard (only in conjunction with Ceramat), loosen
the sensor (hissing sound at the retractable fitting). The dismount guard has a
response time of approx. 2 sec until an error message is released.
!
CAUTION
Only operate the retractable fitting when a sensor is installed!
When the sensor has been removed, it must always be replaced by a dummy!
SensoGate WA 130(X)
Turn the SensoLock ring to ”lock” position to prevent that the retractable
fitting is accidentally moved into the process. This is only possible with the
retractable fitting in “Service“ position (see SensoGate manual).