Configuration
Adaptive Calibration Timer (ACT)
By issuing a Sensoface message, the adaptive calibration timer
reminds you to calibrate the sensor. After expiration of the ACT in-
terval, Sensoface is getting “sad”. Pressing the info key shows the text
“OUT OF CAL TIME CALIBRATE SENSOR” which reminds you that a
calibration is due. The ACT interval is either read automatically from
the sensor settings or can be specified manually (max. 9999 days).
Stressing influences (temperature, measurement in extreme ranges)
shorten the timer interval.
The adaptive cal timer is reset after each calibration.
You can configure the current outputs so that a Sensoface message
generates a 22-mA error signal, see page 81.
Menu item
Action
Choices
Adaptive cal timer
(ACT)
Select using
keys:
AUTO:
The interval stored in
the ISM sensor is used
(default)
MAN:
The interval is specified
manually (0 ... 9999 days)
Press enter to confirm.