Return of products under warranty
Please contact our Service Team before returning a defective device.
Ship the cleaned device to the address you have been given.
If the device has been in contact with process fluids, it must be de-
contaminated/disinfected before shipment. In that case, please attach
a corresponding certificate, for the health and safety of our service
personnel.
Disposal
Please observe the applicable local or national regulations concerning
the disposal of “waste electrical and electronic equipment”.
About this manual:
This manual is intended as a reference guide to your device –
You don't have to read the book from front to back.
Take a look at the Table of Contents or the Index to find the function
you are interested in. Each topic is explained on a double-page spread
with step-by-step instructions on how to configure the desired func-
tion. Clearly legible page numbers and headlines help you to quickly
find the information:
56
57
Configuration
ISM Sensor
Autoclaving counter
Select sensor type
Temperature unit
Temp detection during measurement
(Manual temperature)
Temp detection during calibration
(Manual temperature)
Calibration mode
(AuTo: Buffer set)
ACT - Adaptive calibration timer
TTM - Adaptive maintenance timer
CIP/SIP cycles
Autoclaving counter
Configuration
Autoclaving Counter
After reaching a specified limit value the autoclaving counter gen-
erates a Sensoface message. As soon as the counter has reached
the specified value, Sensoface is getting “sad”. Pressing the info key
shows the text “AuToCLAVE CYCLES oVERRuN” which reminds you
that the maximum number of autoclaving cycles has been reached.
After each autoclaving process, you must manually increment the
autoclaving counter in the SENSoR service menu on the transmitter.
The transmitter displays “INCREMENT AuToCLAVE CYCLE” as confir-
mation. You can configure the current outputs so that a Sensoface
message generates a 22-mA error signal, see page 63.
Menu item
Action
Choices
Autoclaving counter Select using
keys:
oN:
The cycles are specified
manually (0 ... 9999)
Press enter to confirm.
With the autoclaving counter switched on, you must increment the
count after each autoclaving process in the SERVICE menu SENSoR/
AuToCLAVE :
Incrementing the
autoclaving counter
(SERVICE menu)
After having completed
an autoclaving process,
open the SERVICE menu
SENSoR / AuToCLAVE to
increment the autoclav-
ing count.
To do so, select “YES”
and confirm by pressing
enter.
1) Press any arrow key.
2) Select CONF using
keys,
press enter.
3) Select parameter set using
keys,
press enter.
4) Select SENSOR menu using
keys,
press enter.
5) All items of this menu group are indicated by
the “SNS:” code.
Press enter to select menu,
edit using arrow keys (see next page).
Confirm (and proceed) by pressing enter.
6) Exit: Press meas key until the [meas] mode
indicator is displayed.
Left page:
How do I get to
the function
Right page:
Which settings
are provided for
this function
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...