Data Entry of Premeasured Sensors
The DAT calibration mode must have been preset during
configuration.
You can directly enter the values for slope and asymmetry potential
of a sensor. The values must be known, e.g. determined beforehand in
the laboratory.
Display
Action
Remark
Select Calibration.
Press enter to proceed.
“Data Input“
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Enter asymmetry
potential [mV].
Press enter to proceed.
Enter slope [%].
The device displays
the new slope and
asymmetry potential
(at 25 °C).
Sensoface is active.
Use the arrow keys to
select:
• MEAS (exit)
• REPEAT
Press enter to proceed.
Exit:
HOLD is deactivated
with delay.