Calibration is used to adapt the device to the individual sensor
characteristics, namely asymmetry potential and slope.
Access to calibration can be protected with a passcode
(SERVICE menu).
First, you open the calibration menu and select the calibration mode:
CAL_PH
Depending on configuation setting:
AUTO Automatic buffer recognition (Calimatic)
MAN
Manual buffer input
DAT
Input of premeasured electrode data
CAL_ORP
ORP calibration
P_CAL
Product calibration (calibration with sampling)
ISFET-ZERO Zero adjustment. Required for ISFET sensors,
subsequently you can conduct either a
one or a two-point calibration.
CAL_RTD
Temperature probe adjustment
Selecting a Calibration Mode
To preset CAL_PH (CONF menu / configuration):
1) Hold meas key depressed (> 2 s) (measuring mode)
2) Press menu key: the selection menu appears
3) Select CONF mode using left / right arrow key
4) Select “SENSOR“ – “CALMODE“: AUTO, MAN, or DAT.
Press enter to confirm
Selection menu
Selected mode
(blinks)
1
4
3
2
Summary of Contents for Stratos Pro A401B PH
Page 1: ...www knick de Latest Product Information Stratos Pro A401B PH User Manual 094698 ...
Page 11: ...11 Typical Application of Stratos Pro A401B PH IrDA ...
Page 42: ...42 ...
Page 50: ...50 Configuration Select Default Sensor SENSOR ISM for ISM sen sors only Configuration ...
Page 53: ...53 Configuration Configuration Choices Default Relay 1 2 RL1 RL2 Selected in text line ...
Page 104: ...104 ...
Page 148: ...148 ...