![Knick Stratos Pro A401B PH User Manual Download Page 103](http://html.mh-extra.com/html/knick/stratos-pro-a401b-ph/stratos-pro-a401b-ph_user-manual_1980199103.webp)
103
Configuration
Menu item
Action
Choices
Select process
variable
(CHANNEL)
Select desired process
variable using
keys.
Press enter to confirm.
Limit 2 function
(FUNCTION)
Select desired function
using arrow keys.
Press enter to confirm.
Limit 2 icon:
Limit 2 contact type
(CONTACT)
N/O: normally open
contact
N/C: normally closed
contact
Select using
keys.
Press enter to confirm.
Limit 2 setpoint
(LEVEL)
Enter setpoint using
keys.
Press enter to confirm.
Limit 2 hysteresis
(HYSTERESIS)
Select hysteresis using
keys.
Press enter to confirm.
Limit 2 delay
(DELAYTIME)
The contact is activated
with delay (deactivated
without delay)
Adjust delay using
keys.
Press enter to confirm.
Limit Lo
Hyst
Setpoint
Contact
0
1
Signal
Limit Hi
Hysteresis -
Signal
Setpoint
Contact
0
1
5
Application of hysteresis: