81
Configuration
Menu item
Action
Choices
Delay
Enter value using
keys.
Press enter to confirm.
0...600 SEC
(010 SEC)
Sensocheck
Select Sensocheck
(continuous monitoring
of sensor).
Select ON or OFF using
keys.
Press enter to confirm
(At the same time,
Sensoface is activated.
With OFF, Sensoface is
also switched off.)
ON/OFF
5
Alarm contact
The alarm contact is closed during normal
operation (N/C).
It opens in the case of alarm or power outage.
As a result, a failure message is provided even
in the case of line breakage (fail-safe behavior).
For contact ratings, see Specifications.
17
18
Alarm
R3
Error messages can also be signaled by a 22 mA output current
(see Error messages and Configuration Output 1/Output 2).
Operating behavior of the alarm contact: see Operating States table.
The alarm delay time delays the color change of the display back-
lighting to red, the 22 mA signal (if configured), and the alarm contact
switching.