Error Messages
Error
Info text
(is displayed in case of
fault when the Info key is
pressed)
Problem
Possible causes
ERR 99
DEVICE FAILURE
Error in factory settings
EEPROM or RAM defective
This error message only occurs
in the case of a total defect. The
device must be repaired and
recalibrated at the factory.
ERR 98
CONFIGURATION ERROR
Error in configuration or
calibration data
Memory error in device program
Configuration or calibration data
defective; completely reconfig-
ure and recalibrate the device.
ERR 97
NO MODULE INSTALLED
No module
Please have the module
replaced in the factory.
ERR 96
WRONG MODULE
Wrong module
Please have the module replaced
in the factory.
ERR 95
SYSTEM ERROR
System error
Restart required.
If error still persists, send in the
device for repair.
ERR 100 INVALID SPAN OUT1
Span Out1 configuration error
ERR 101 INVALID SPAN OUT2
Span Out2 configuration error
ERR 104 INVALID PARAMETER
CONTROLLER
Controller configuration error
ERR 105 INVALID SPAN I-INPUT
I-Input configuration error