127
Specifications
CONDI input
Input for toroidal (electrodeless) conductivity sensors
SE 655, SE 656, SE 660, SE 670
Effective ranges
Conductivity
0.000 ... 1999 mS/cm
Concentration
0.00 ... 100.0% by wt
Salinity
0.0 ... 45.0 % (0 ... 35 °C / 32 ... 95 °F)
Display ranges
Conductivity
0.000 ... 9.999 mS/cm (not with SE 660 / SE 670)
00.00 ... 99.99 mS/cm
000.0 ... 999.9 mS/cm
0000 ... 1999 mS/cm
0.000 ... 9.999 S/cm
00.00 ... 99.99 S/cm
Concentration
0.00 ... 9.99 % / 10.0 ... +100 %
Salinity
0.0 ... 45.0 % (0 ... 35 °C / 32 ... 95 °F)
Response (T90)
Approx. 1 s
Meas. error
1,2,3)
< 1% meas. val. + 0.005 mS
Temperature
compensation
*
(OFF)
Without
(LIN)
Linear characteristic 00.00 ... 19.99%/K
(reference temp user-defined)
(NLF)
Natural waters to EN 27888
(reference temp 25 °C / 77 °F)
(NACL)
NaCl from 0 to 26 % by wt (0 ... 120 °C / 32 ... 248 °F)
(reference temp 25 °C / 77 °F)
Concentration
determination*
–01– NaCl
0 – 26 wt% (0 °C /32 °F) ... 0 – 28 wt% (100 °C/212 °F)
–02– HCl
0 – 18 wt% (-20 °C /-4 °F) ... 0 – 18 wt% (50 °C/122 °F)
–03– NaOH
0 – 13 wt% (0 °C /32 °F) ... 0 – 24 wt% (100 °C/212 °F)
–04– H
2
SO
4
0 – 26 wt% (-17 °C/1.4 °F)…0 – 37 wt% (110 °C/230 °F)
–05– HNO
3
0 – 30 wt% (-20 °C /-4 °F) ... 0 – 30 wt% (50 °C/122 °F)
–06– H
2
SO
4
94 – 99 wt% (-17 °C/1.4 °F)…89 – 99 wt% (115 °C/239 °F)
–07– HCl
22 – 39 wt% (-20 °C/-4 °F)…22 – 39 wt% (50 °C/122 °F)
–08– HNO
3
35 – 96 wt% (-20 °C/-4 °F)…35 – 96 wt% (50 °C/122 °F)
–09– H
2
SO
4
28 – 88 wt% (-17 °C/1.4 °F)…39 – 88 wt% (115 °C/239 °F)
–10– NaOH
15 – 50 wt% (0 °C/32 °F)…35 – 50 wt% (100 °C/212 °F)
Summary of Contents for Stratos Pro A401B CONDI
Page 1: ...Stratos Pro A401B CONDI User Manual www knick de Latest Product Information 094711 ...
Page 11: ...11 Typical Application of Stratos Pro A401B CONDI ...
Page 90: ...90 ...
Page 102: ...102 ...
Page 113: ...113 ...
Page 123: ...123 ...
Page 143: ...143 ...