108
Calibration
Calibration by Input of Cell Factor
You can directly enter the value for the cell factor of a sensor. This
value must be known, e.g. determined beforehand in the laboratory.
The selected process variable and the temperature are displayed.
This method is suitable for all process variables.
Display
Action
Remark
Select Calibration.
Press enter to proceed.
Select CAL_CELL
calibration method.
Press enter to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Enter cell factor.
Press enter to proceed.
The selected process
variable and the
temperature are
displayed.
The device shows the
calculated cell factor
and zero point
(at 25 °C).
Sensoface is active.
Use the arrow keys to
select:
• MEAS (exit)
• REPEAT
Press enter to proceed.
Exit:
HOLD is deactivated
after a short time.
Please refer to the Specifications for the nominal cell factor. When
measuring in a restricted space, the individual cell factor must be
determined.