88
Digital Sensors
Replacing a Sensor
A digital sensor should only be replaced during HOLD mode to pre
vent unintended reactions of the outputs or contacts. When you first
want to calibrate the new sensor, it can also be replaced in calibration
mode.
Step
Action/Display
Remark
Select HOLD mode
Press
menu
key to
call the selection
menu,
select HOLD using
the
keys, press
enter
to
confirm.
Now the device is in
HOLD mode. The HOLD
mode can also be acti
vated externally via the
HOLD input.
During HOLD the output
current is frozen at its
last value or set to a fixed
value.
Disconnect and
remove old sensor
Install and connect
new sensor.
Temporary messages
which are activated dur
ing the replacement are
indicated but not output
to the alarm contact and
not entered in the log
book.
Wait until the sensor
data are displayed.
MEMO SENS