Configuration
Parameter Set Selection
Note:
Manual selection of parameter sets must have been preset in
the CONFIG > PARSET menu. Default setting is a fixed parameter set A.
Wrong settings change the measurement properties!
Manual switchover of parameter sets A/B
Display
Action
To switch between parameter sets:
Press
meas
.
PARSET blinks in the lower line.
Select parameter set using
and
keys
Press
enter
to confirm.
Cancel by pressing
meas
.
External switchover of parameter sets A/B (TAN SW-A005)
You can switch between parameter sets A and B by applying a signal
to the CONTROL input (parameter setting: CONTR‑IN > PARSET).
Parameter set A active
0...2 V AC/DC
Parameter set B active
10...30 V AC/DC
e.g., process control system
Max. 30 V AC/DC
CONTROL
input
13
14
Stratos Pro A201