![Knick Stratos Pro A201COND User Manual Download Page 27](http://html.mh-extra.com/html/knick/stratos-pro-a201cond/stratos-pro-a201cond_user-manual_1980116027.webp)
27
Sensor
Example 8:
Measuring task:
Conductivity, temperature
Sensor:
Memosens
Wiring Examples
W
hit
e
Br
own
Yello
w
Shield
Gr
een
Cable
De
vic
e
The Memosens sensor is connected to the RS-485 interface of the
device – for an A2... Series (2-wire) device, the measuring module
slot must be empty. Therefore, first remove the measuring module
from the slot (see next page). Open the CONFIG menu and select
“MEMOSENS”. The connected Memosens sensor is then automatically
recognized during start-up of the transmitter.