Electrical Installation
Before commencing with the installation, make sure that all lines to be
connected are de-energized.
Observe the safety instructions; see page 7.
Cable Glands
In a hazardous location, only cable glands with suitable approvals may
be used. The installation instructions of the manufacturer must be
observed.
Cable glands
5 cable glands M20 x 1.5 A/F 24 mm
WISKA type ESKE/1 M20
Clamping ranges
Standard sealing insert: 7 … 13 mm
Reduction sealing insert: 4 … 8 mm
Multiple sealing insert: 5.85 … 6.5 mm
Tensile strain
Not permitted; Only suitable for “fixed installation”
CAUTION!
Risk of losing the specified ingress protection.
Fasten the cable glands and screw together the housing correctly.
Observe the permissible cable diameters and tightening torques.
Only use original accessories and spare parts.
NOTICE!
Strip the insulation from the wires using a suitable tool to
prevent damage. For stripping length, see Specifications.
1) Wire the current outputs. Deactivate unused current outputs in the
parameter settings or use jumpers.
2) Wire the inputs as necessary.
3) Connect the sensor.
4) Check whether all connections are correctly wired.
5) Close the housing and successively tighten the enclosure screws in a
diagonal pattern.