67
Configuration
Autoclaving Counter
After reaching a specified limit value the autoclaving counter gen‑
erates a Sensoface message. As soon as the counter has reached
the specified value, Sensoface is getting “sad”. Pressing the info key
shows the text “AUTOCLAVE CYCLES OVERRUN” which reminds you
that the maximum number of autoclaving cycles has been reached.
After each autoclaving process, you must manually increment the
autoclaving counter in the SENSOR service menu on the transmitter.
The transmitter displays “INCREMENT AUTOCLAVE CYCLE” as confir‑
mation. You can configure the current outputs so that a Sensoface
message generates a 22‑mA error signal, see page 75.
Menu item
Action
Choices
Autoclaving counter
Select using
keys:
ON:
The cycles are specified
manually (0 ... 9999)
Press
enter
to confirm.
With the autoclaving counter switched on, you must increment the
count after each autoclaving process in the SERVICE menu SENSOR/
AUTOCLAVE :
Incrementing the
autoclaving counter
(SERVICE menu)
After having completed
an autoclaving process,
open the SERVICE menu
SENSOR / AUTOCLAVE to
increment the autoclav‑
ing count.
To do so, select “
YES
”
and confirm by pressing
enter
.