151
Error Messages
Error
Info text
(is displayed in case of
fault when the Info key is
pressed)
Problem
Possible causes
ERR 99
DEVICE FAILURE
Error in factory settings
EEPROM or RAM defective
This error message only occurs
in the case of a total defect. The
device must be repaired and
recalibrated at the factory.
ERR 98
CONFIGURATION ERROR
Error in configuration or
calibration data
Memory error in device program
Configuration or calibration data
defective; completely reconfig-
ure and recalibrate the device.
ERR 97
NO MODULE INSTALLED
No module
Please have the module installed
in the factory.
ERR 96
WRONG MODULE
Wrong module
Please have the module replaced
in the factory.
ERR 95
SYSTEM ERROR
System error
Restart required.
If error still persists, send in the
device for repair.
ERR 01
NO SENSOR
pH sensor *
The sensor is not recognized:
Check connections.
Check cables/sensor.
Replace as required.
ERR 02
WRONG SENSOR
Wrong sensor *
Replace the sensor.
ERR 04
SENSOR FAILURE
Failure in sensor *
Replace the sensor.