![Knick Stratos Pro A201 PH User Manual Download Page 109](http://html.mh-extra.com/html/knick/stratos-pro-a201-ph/stratos-pro-a201-ph_user-manual_1980194109.webp)
109
Converting Slope to mV
Converting slope [%] to slope [mV/pH] at 25 °C
%
mV/pH
78
46.2
80
47.4
82
48.5
84
49.7
86
50.9
88
52.1
90
53.3
92
54.5
94
55.6
96
56.8
98
58.0
100 59.2
102 60.4
Converting asymmetry potential to sensor zero point
= Sensor zero
= Asymmetry potential
= Slope
ZERO = 7 -