Configuration
Time and Date
Control of the calibration and cleaning cycles is based on the time and date of the
integrated real-time clock.
In measuring mode the time is shown in the lower display. When using digital
sensors, the calibration data is written in the sensor head. In addition, the logbook
entries (cf Diagnostics) are provided with a time stamp.
Note:
There is no automatic switchover from winter to summer time!
Be sure to manually adjust the time!
Sensor Verification (TAG, GROUP)
When Memosens sensors are calibrated in the lab, it is often useful and sometimes
even mandatory that these sensors will be operated again at the same measuring
points or at a defined group of measuring points. To ensure this, you can save the
respective measuring point (TAG) or group of measuring points (GROUP) in the
sensor. TAG and GROUP can be specified by the calibration tool or automatically
entered by the transmitter. When connecting an MS sensor to the transmitter, it can
be checked if the sensor contains the correct TAG or belongs to the correct GROUP.
If not, a message will be generated, Sensoface gets “sad”, and the display backlight-
ing turns purple (magenta). The “sad” Sensoface icon can also be signaled by a
22 mA error current. Sensor verification can be switched on in the Configuration in
two steps as TAG and GROUP if required.
When no measuring point or group of measuring points is saved in the sensor, e.g.,
when using a new sensor, Stratos enters its own TAG and GROUP. When sensor veri-
fication is switched off, Stratos always enters its own measuring point and group.
A possibly existing TAG/GROUP will be overwritten.
Menu item
Action
Choices
Measuring point
(TAG)
In the lower display line you can
enter a designation for the measuring
point (TAG) and for a group of mea-
suring points (GROUP) if applicable.
Up to 32 digits are possible.
By pressing meas (repeatedly) in the
measuring mode you can view the
tag number.
Select character using
keys,
select next digit using
keys.
Press enter to confirm.
A...Z, 0...9, – +
< > ? / @
The first 10
characters are
seen in the
display with-
out scrolling.