Configuration
Menu item
Action
Choices
Calibration timer
Adjust CALTIMER using
keys:
OFF: No timer
ADAPT: Maximum cal
cycle (adjust in the next
step)
FIX: Fixed cal cycle (adjust
in the next step)
Press enter to confirm.
With ADAPT, the calibra-
tion cycle is automatically
reduced depending on
the sensor load (high
temperatures and pH
values) and for digital
sensors also depending
on the sensor wear
Calibration cycle
Only with FIX/ADAPT:
Modify digit using
keys,
select next digit using
keys.
Press enter to confirm.
5
Note for the calibration timer:
When Sensocheck has been activated in the Configuration – Alarm
menu, the expiration of the calibration interval is indicated by
Sensoface:
Display
Status
+
+
Over 80 % of the calibration interval has
already past.
The calibration interval has been exceeded.
The calibration timer settings apply to both parameter sets A and B.
The time remaining until the next due calibration can be seen in the
diagnostics menu (see Diagnostics chapter).