33
Configuration
Parameter set A/B: configurable menu groups
The device provides 2 parameter sets “A“ and “B“. By switching between the param-
eter sets you can adapt the device to different measurement situations, for example.
Parameter set “B“ only permits setting of process-related parameters.
Menu group
Parameter set A
Parameter set B
Sensor selection
---
Current output 1
Current output 1
Current output 2
Current output 2
Compensation
Compensation
Control input
---
Alarm mode
Alarm mode
Parameter set selection
---
Setting the clock
---
TAG
TAG of measuring point
GROUP
GROUP of measuring
points
External switchover of parameter sets A/B
You can switch between parameter sets A and B by applying a signal
to the CONTROL input (parameter setting:
).
Parameter set A active
0...2 V AC/DC
Parameter set B active
10...30 V AC/DC
e.g. process control system
Max. 30 V AC/DC
CONTROL
input
13
14
Stratos Pro A2...