Sensoface
Display Problem
Status
Sensor wear
(for digital
sensors only)
High temperatures and pH
values have caused a wear of
over 80%.
The sensor should be replaced
soon.
Wear is at 100%.
Replace sensor.
SENSOR WEAR
CHANGE SENSOR (DLI)
Replace sensor
AUTOCLAVE CYCLES
OVERRUN
Maximally permitted number of auto-
claving cycles has been reached.
Replace sensor or increment autoclaving
counter.
SIP CYCLES OVERRUN
Maximally permitted number of
sterilizing cycles has been reached.
Replace sensor or increment SIP counter.
CIP CYCLES OVERRUN
Maximally permitted number of
cleaning cycles has been reached.
Replace sensor or increment CIP
counter.