102
Menu item
Remark
Display example:
Displaying currently measured values
(sensor monitor) with HOLD mode activated:
Select MONITOR using
, press enter to confirm.
Select variable in the bottom text line using
.
The selected parameter is shown in the upper dis-
play line.
As the device is in HOLD mode, you can perform
validations using simulators without influencing the
signal outputs.
Return to Service menu:
Hold meas depressed for longer than 2 sec.
Press meas once more to return to measurement.
Resetting the adaptive maintenance timer
Here, the interval is reset to its initial value. To do so,
select “TTM RESET = YES“ and confirm by pressing
enter.
Incrementing the autoclaving counter
After having completed an autoclaving process,
you must increment the autoclaving count.
To do so, select “YES” and confirm by pressing enter.
The device confirms with
“INCREMENT AUTOCLAVE CYCLE“.
Specifying the current at outputs 1 and 2:
Select OUT1 or OUT2 using the
keys,
press enter to confirm.
Enter a valid current value for the respective output
using
keys.
Press enter to confirm.
For checking purposes, the actual output current is
shown in the bottom right corner of the display.
End by pressing enter or meas.