Stratos Multi E401N
213
13.4.8 Firmware Update (FW-E106)
Note:
First check whether your device really requires a firmware update.
The FW-E106 add-on function must be activated in the device with a TAN for the firmware update.
Stratos Multi features a standard microcontroller and a microcontroller for communication.
The firmware of both microcontrollers can be updated. The firmware files are named as follows:
• Standard microcontroller: FW: xx.xx.xx. Build xxxxx
• Communication microcontroller: IF-4000: xx.xx.xx. Build xxxxx
The two updates may have to be carried out one after the other.
NOTICE!
The device is not operable during a firmware update.
Its outputs are in an undefined state.
After a firmware update, the configuration must be checked.
Note:
Before the firmware update, we recommend saving the previous version on the FW Update
Card.
Firmware Update with FW Update Card
WARNING! Shock potential.
When opening the device, there may be dangerous touch voltages in
the terminal compartment. Professional installation guarantees direct contact protection.
The memory card can be replaced during operation. When doing so, maintain sufficient distance from
the mains connection cables and do not use tools.
01.
Open the housing.
02.
Insert the FW Update Card into the memory card slot in the front unit.
✓
The FW Update Card icon appears on the display
.
03.
Close the housing.
04.
If required, save the firmware currently installed on the device:
Menu
Parameter Setting
System Control
Firmware Update
Firmware Backup
05.
Start the backup with the
right softkey: Start
.
✓
When the backup has finished, the device returns to measuring mode.
06.
Load the firmware update:
Menu
Parameter Setting
System Control
Firmware Update
Update Firmware
07.
Select a version using the arrow keys.
08.
Confirm with
enter
.
09.
Start the firmware update with the
right softkey: Start
.
✓
Bootloader starts: deletes memory, overwrites memory with new firmware, and recalculates it.
✓
The device returns to measuring mode.
10.
As required, carry out the second firmware update (IF-4000).
11.
When the update has finished, open the housing and remove the FW Update Card.
12.
Close and screw the housing together.
13.
Check the configuration.