Inserting a Module
Measuring modules for connection of analog conductivity sensors:
Measuring modules for the connection of analog sensors are simply inserted into
the module slot. Upon initial start-up, the analyzer automatically recognizes the
module and adjusts the software correspondingly. When you replace the measur-
ing module, you must select the corresponding measuring function in the “Service”
menu.
This does not apply to the multi-channel module for dual conductivity measurement
and to the connection of Memosens sensors. Here, you will be prompted to select
the desired measuring function upon first start-up.
Multi-channel module for connection of analog sensors:
dual conductivity
For this module, you must select the operating mode (“device type”) in the
configuration menu.
Changing the Measuring Function
In the “Service” menu you can select another measuring function at any time.