![Knick Stratos Evo A402 User Manual Download Page 140](http://html.mh-extra.com/html/knick/stratos-evo-a402/stratos-evo-a402_user-manual_1980181140.webp)
140
Calibration by input of cell factor (toroidal sensors)
You can directly enter the value for the cell factor of a sensor. The value must be
known, eg, determined beforehand in the laboratory. The selected process variable
and the temperature are displayed. This method is suitable for all process variables.
Display
Action
Remark
Select Calibration.
Press enter to proceed.
Select CAL_CELL
calibration method.
Press enter to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Enter cell factor.
Press enter to proceed.
The selected process
variable and the tem-
perature are displayed.
The device shows the cal-
culated cell factor and zero
point (at 25 °C).
Sensoface is active.
Use the arrow keys to
select:
• MEAS (exit)
• REPEAT
Press enter to proceed.
Exit:
HOLD is deactivated
after a short time.
Please refer to the Specifications for the nominal cell factor.
When measuring in a restricted space, the individual cell factor must be determined.
Calibration by Input of Cell Factor
Summary of Contents for Stratos Evo A402
Page 2: ......
Page 70: ...70 70 ...
Page 120: ...120 ...
Page 132: ...132 ...
Page 142: ...142 ...
Page 175: ...175 Specifications ...
Page 194: ...194 Notes ...
Page 195: ...195 Notes ...