In the alarm menu you can configure flow monitoring.
When you have set CONTROL to FLOW, you can specify 2 additional limit values for
maximum and minimum flow. If the measured value lies outside this range, an alarm
message and a 22-mA error signal (if configured) will be generated.
Menu item
Action
Select
Select function of
CONTROL input
Select using
keys, confirm
by pressing enter
(selecting parameter set A/B via
signal at CONTROL input)
(for connecting a pulse-output
flow meter)
Adjust to flow meter:
With “Flow” selected, you must
adjust the device to the flow
meter used.
Enter value using arrow keys,
confirm by pressing enter
12000 pulses/liter
Display
Flow measurement in measuring mode
Display
Flow measurement (sensor monitor)
Configuring the CONTROL Input
Note: The response speed may be reduced because the values are averaged.