Configuring the Current Output
HOLD active
21
4
HOLD active
Output current
[mA]
Output signal for HOLD
FIX setting = 21.0 mA
Output signal for HOLD
LAST setting
Output signal during HOLD:
Menu item
Action
Choices
Output current during
error message
The output current can be set
to 22 mA in the case of error
messages.
Select ON or OFF using
.
Press enter to confirm.
Output current during
Sensoface messages
The output current can be set to
22 mA in the case of Sensoface
messages.
Select ON or OFF using
.
Press enter to confirm.
Output current during
HOLD
LAST: During HOLD the last
measured value is maintained at
the output.
FIX: During HOLD a value (to be
entered) is maintained at the
output.
Select using
.
Press enter to confirm.
Output current for HOLD
FIX
Only with FIX selected:
Enter current which is to flow at
the output during HOLD
Enter value using
.
Press enter to confirm.