30
Calibration Chamber
SensoGate WA 132
1) Remove the outlet connector
(AM)
. Take off the loose flange if
required.
2) Loosen and remove the two
screws
(AK)
from the calibra-
tion chamber (using screwdriver
TX 25).
K
P
N
3) Pull the calibration chamber
(K)
vertically out of the process
adaptation
(P)
.
4) Push the flange bushing
(N)
downwards out of the process
adaptation. Now, the gaskets are
accessible and can be checked
and replaced if required.
AK
AK
Removing the Calibration Chamber
AM