Installing the Outlet Hose
SensoGate® WA 131 H
Push the outlet hose
(A)
over the connection nipple
(C)
. Then secure the the hose connection with pipe
clamp
(B)
.
The outlet hose serves for draining the rinse or
calibration solutions.
Specifications
Thread: G 1/8 internal, outlet port
Length: 3 meters
Nominal width: DN 8
Materials: Hose
EPDM
Connection
nipple
PVDF
O-ring
8
x
1.5
FKM
(Viton)
A
C
B
Also when you have a sensor lock-gate without rinse connection, you should install the included
outlet hose because of the applied process pressure.
When the sensor moves into and out of the PROCESS position, pressure is build up in the fitting.
Pressurized process medium that might have intruded is drained through the outlet hose.