9
ISM
Plug and Measure
An ISM sensor is immediately identified after being connected:
ISM sensor identified
25.6 °C
Proceed
Manufacturer:
Article No.:
99.2 % Air
Adjustment:
Serial number:
Sensor:
Mettler Toledo
52002559
09/06/10 08:15
0000313
InPro 6900
Date 09/06/10
Parameter set
Message list 1 messg.
Return
Warn New sensor, adjustment required
25.6 °C
99.2 % Air
Date 09/06/10
Parameter set
Failure Message
(Incorrect Measured Values)
Measured value, alarm icon, and module slot
identifier are flashing.
The flashing means:
NOTICE!
The displayed value is no “valid” measured value!
All sensor-typical parameters are automatically
sent to the analyzer.
These are, for example, the measuring range,
zero and slope of the sensor, but also the type
of temperature probe. Without any further
parameter setting, measurement starts at once,
the measuring temperature is simultaneously
detected.
Premeasured ISM sensors can be used for mea-
surement without previous calibration.
The ISM logo is displayed as long as an ISM
sensor is connected.
When the ISM sensor has not been adjusted, the
“maintenance request” icon is displayed.
A new entry is added to the message list of the
Diagnostics menu:
Warn New sensor, adjustment required
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.