31
Menu Display
Zero correction
Select “Module OXY 3400-067”
The module is in function check
(HOLD) mode. The assigned current
outputs and relay contacts behave as
configured (BASE).
Press
enter
to confirm.
Zero correction:
Display of measured sensor current.
• Enter input current
for zero point
Confirm with “OK”.
Select “Zero correction”
calibration method
Press
enter
to confirm.
Zero Correction
The sensor models SE 7*6 ... and SE7*7 ... have a very low zero current.
For trace measurements below 500 ppb, the zero point should be calibrated.
If a zero correction is performed, the sensor should remain for at least 10 to
60 minutes in the calibration medium (media containing CO
2
at least 120 min)
to obtain stable, non-drifting values. During zero correction, a drift check is not
performed.
Calibration
Return
25.6°C
Module PH 3400-032
Module CONDI 3400-051
Info
80.3%Air
Module OXY 3400-067
Module OXY 3400-067
Return
25.6°C
80.3%Air
Automatic - Water
Automatic - Air
Product calibration Sat
Data entry
Zero correction
Zero correction
Direct displacement
of sensor zero point
25.6 °C
80.3%Air
Measured sensor current
Abort
-0.009nA
Zero
+
0
.000nA
OK
Calibration / Adjustment
Zero Correction
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.