Set rinse contact parameters
• Set rinse interval
• Set rinse duration
• During the defined "lead time" the
"HOLD" mode is active.
• Select contact type
(e.g. "N/O")
Rinse Contact
Parameter setting/Module BASE/Relay contacts/Usage/Rinse contact
Menu Display
Configuring the rinse contact
Relay contacts, usage
• Open parameter setting
• Enter passcode
• Select “Module BASE”
• Select contact e.g. K1)
• “Rinse contact” (Fig.)
Contact K1 (Administrator)
19.2°C
Return
Hysteresis
Variable
Limit value
7.00 pH
Usage
Effective direction
Contact type
Limit value
NAMUR maintenance
NAMUR HOLD
Rinse contact
Parameter set B active
USP output
Rinse contact
Contact K1 (Administrator)
19.2°C
Return
Rinse duration
Rinse interval
Rinsing lead time
7.00pH
Usage
Measurement lead time
Contact type
0010 s
Rinse contact
000.0 h
0010 s
N/O
0016 s
Please note when configuring the "Rinse contact" function
• "HOLD" mode (e.g. during parameter setting) delays the execution of the
"Rinse contact" function.
• Up to 3 rinse functions (contacts K1 ... K3) can be configured independently.
• The individual rinse functions are not synchronized with each other.
Time response
HOLD
Contact Kx
Lead time
Rinse duration
Shut-off delay