Zero correction
Calibration / Adjustment
Zero Correction
Menu Display
Zero correction
Select module: CONDI
Confirm with enter
Permissible zero point deviation
depends on the sensor type. For the
SE 655 (SE656), it is ± 0.050 mS/cm.
Press Adjust to take over the
calibration data.
Select calibration method
“Zero correction”
Remove the sensor from the process
and dry it.
Confirm with enter
Zero Correction
Adjustment of zero point / Automatic determination of the zero point in air
Every electrodeless (toroidal) conductivity sensor has its individual zero point.
When measuring low conductivity values, accuracy can be increased by
adjusting the zero point.
During calibration the module is in HOLD mode.
Current outputs and relay contacts of the module behave as configured
(Module BASE).