Adjustment
Adjustment
means that the cell factor determined by a calibration is taken over.
It is entered in the calibration record. (Cal record can be called in the
Diagnostics menu for the CONDI 3400(X)-051 module.)
The value is only effective for calculating the measured variables when the
calibration has been terminated with an adjustment.
A passcode ensures that an adjustment can only be performed by an autho-
rized person (Administrator).
The Operator can check the current sensor data by a calibration and inform the
Administrator when there are deviations.
You can use the additional function SW 3400-107 for granting access rights
(passcodes) and for Audit Trail (continuous data recording and backup
according to FDA 21 CFR Part 11).
Operator
(without administrator rights)
After calibration, change to measuring
mode. Inform Administrator.
When opening the menu (Calibration,
respective module), the Administrator
sees all data of the last calibration and
can take over the values or perform a
new calibration.
Calibration data record
Cal mode
25.6 °C
1.225 mS/cm
Conductivity
Automatic
0249 mS/cm
End
Adjust
Calibration
Cal temp
03/31/10 12:30
25.6 °C
Menu Display
Adjustment after calibration
Administrator
With the corresponding access rights,
the device can immediately be adjust-
ed after calibration. The calibration
values are taken over for calculating
the measured variables.
Module CONDI 3400-051
Return
25.6°C
View/adjust calibration data record
1.225 mS/cm
Start new calibration
Stored calibration data record
Calibration 03/31/10 12:30
Cell factor
0.998 S/cm
Adjustment