78
Display text
Time [s]
01: Probe in SERVICE
02: Cleaning agent
0020 s
| Measure -> Service
03: Rinse water ON
0060 s
|
04: Rinse water OFF
0002 s
|
05: Purge air ON
0005 s
|
06: Purge air OFF
0002 s
|
07: Wait position
| Service position
08: Rinse water ON
0010 s
|
09: Rinse water OFF
0002 s
|
10: Probe in MEASURE
0005 s
| Service -> Measuring
11: Program end
Parameter Setting: Program Flows
Service
Program step
Probe: Measure
Probe: Service
Rinse water
HOLD
DCS ML3
(Program running)
01
02
03
04
05
06
Message (display)
Purge air
(AUX1 valve)
Part 1: Measure —> SERVICE
Cleaning agent
(Pump III)
Stroke 1 Stroke 2
Exposure time
07
08
09
10
11
Part 2: SERVICE —> Measure
W
ait p
osition
Service can also be is started via a DCS input signal at input M/S
of the Unical 9000 probe controller.
Service program