• Select Curve,
e.g. “linear”: The measured variable is
represented by a linear output cur-
rent curve. The desired range of the
measured variable is specified by the
values for “Start” and “End”.
Current Outputs, Contacts, OK Inputs
Select menu: Parameter setting/Module BASE
Note: HOLD mode active
Menu Display
Parameter setting
BASE module
• Select measured variable
Output current
[pH]
20
4
14
0
Assignment of Measured Values: Start (4 mA) and End (20 mA)
[pH]
20
4
7
5
[mA]
Example 1: Range pH 0 - 14
Example 2: Range pH 5 - 7
Advantage: Higher resolution in range of interest
[mA]
To configure current output
• Call up parameter setting
• Enter passcode
• Select “Module BASE”
• Select “Output current I1”
Configuring a Current Output
• Call up parameter setting
• Enter passcode
• Select “Module BASE”
• Select “Output current ...”
Module BASE (Administrator)
19.2°C
Return
Contact K2 (NAMUR HOLD)
Contact K4 (NAMUR Failure)
Contact K3 (NAMUR maintenance)
7.00 pH
Output current I2
Contact K1 (Limit)
Inputs OK1, OK2
Block
Output current I1 (Administrator)
OK
19.0 °C
7.00 pH
Variable
Output
Abort
Curve
Linear
°C
Trilinear
Function
Table
Output current I1 (Administrator)
OK
19.0 °C
7.00 pH
Curve
Output
Behavior during messages
Start
Abort
Variable
°C
rH
S/cm
°C
g/kg
Off
pH
End
Start
End
Output filter
5 7
Current outputs
BASE module
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...