208
Icon
Explanation of icons important for the
PHU 3400(X)-110 module
The analyzer is in measuring mode.
The analyzer is in calibration mode. HOLD mode active for currently calibrated module.
The analyzer is in maintenance mode. HOLD mode active.
The analyzer is in parameter setting mode. HOLD mode active.
The analyzer is in diagnostics mode.
HOLD. The NAMUR “HOLD” contact is active (factory setting: Module BASE, Contact K2,
N/O contact). Current outputs as configured:
• Current meas.:
The currently measured value appears at the current output
• Last usable value: The last measured value is held at the current output
• Fixed 22 mA:
The output current is at 22 mA
Calibration is performed (progress display)
Calibration - Step 1 of product calibration has been executed.
The analyzer is waiting for the sample values.
Temperature by manual input
In the plaintext display in front of a menu line:
Access to next menu level with enter
In the plaintext display in front of a menu line when it has been blocked by the
Administrator against access from the Operator level.
Designates the module slot (1, 2, or 3), allowing the clear assignment of measured-value/
parameter displays in the case of identical module types.
Indicates the active parameter set .(The analyzer provides two parameter sets A and B.
Up to 5 sets can be added using additional functions and SmartMedia card.)
Failure. The NAMUR “failure” contact is active (factory setting: Module BASE, Contact K4,
N/C contact). To view error message, call up: Diagnostics menu/Message list
Maintenance. The NAMUR “maintenance request” contact is active (factory setting: Module
BASE, Contact K2, N/O contact). To view error message, call: Diagnostics menu/Message list
NAMUR signals
Temperature compensation for process medium is active (Linear/Ultrapure water/Table)
TC
Delta function is active (Output value = measured value – delta value)
∆
man
Overview of icons
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...