141
Menu Display
Diagnostics functions
Open diagnostics
From the measuring mode:
Press menu key to select menu.
Select diagnostics using arrow keys,
confirm with enter.
The “Diagnostics” menu gives an
overview of all functions available.
Functions which have been set as
“Favorite” can be directly accessed
from the measuring mode.
Diagnostics Functions
General status information of the measuring system
Select menu: Diagnostics - Logbook
Logbook
The last 50 events are recorded with
message identifier, date, time, and
module concerned.
This permits quality management
documentation to ISO 9000 et seq.
Extended logbook: SmartMedia card
(SW 3400-104)
Menu selection
Return to meas
7.00 pH
25.6°C
Select:
[enter]
Lingua
Diagnostics
Return
25.6°C
Set favorite
7.00 pH
Logbook
Message list
Device description
Module FRONT
Module BASE
Diagnostics
Point of meas description
Return
25.6°C
7.00 pH
Note:
Tank_2
Measurement point:
Point of Meas Description
Allows entering a tag number and a
note. Select position: left/right arrow
key, select character: up/down arrow
key. Confirm the entry with enter.
04/03/2007 smith
Point of meas description
Logbook
Return
U229 04/13/07 09:50 • Warn Sensor dismo
F224 04/12/07 17:52 Measurement active
F222 12.04.07 17:44 Parameter setting active
F222 04/13/07 09:36 Parameter setting activ
25.6 °C
7.00 pH
U229 11.04.07 06:51 Warn Sensor dismou
U229 11.04.07 06:50 • Warn Sensor dismou
Summary of Contents for Protos 3400
Page 12: ...12 ...
Page 160: ...160 ...
Page 177: ...177 Appendix Buffer Table Mettler Toledo ...
Page 179: ...179 Appendix Buffer Table DIN 19267 ...
Page 181: ...181 Appendix Buffer Table Techn Buffers to NIST ...
Page 182: ...182 Appendix Buffer Table Hamilton ...
Page 183: ...183 Appendix Buffer Table Kraft ...
Page 184: ...184 Appendix Buffer Table Hamilton A ...
Page 185: ...185 Appendix Buffer Table Hamilton B ...
Page 186: ...186 Appendix Buffer Table HACH ...
Page 206: ...206 Index ...
Page 207: ...207 Index ...