Parameter Setting: Program Flows
Rinsing (Continuous)
Display text
Time [s]
01: Probe in SERVICE
02: Rinse water ON
0060 sec
03: Rinse water OFF
0002 sec
04: Probe in MEASURE
0005 sec
05: Program end
Program step
Probe: Measure
Probe: Service
Rinse water
HOLD
DCS ML3
(Program running)
01
02
03
04
05
t
[s]
Cleaning (continuous) can also be started via a DCS input signal at input BIN1
of the Uniclean 900 probe controller.
Message (display)
Parameter setting: Program
flows
Rinse program