Options
Enabling the Option 001 SOP
1) In measuring mode, press the Menu softkey.
2) Select “Configuration” and confirm by pressing enter.
3) Select Option “001 SOP” and enter your activation code.
Adapt SOP Cal
Here, you specify which buffers are to be used in which sequence.
You can combine buffer solutions from different buffer sets. Please note that the
minimum distance allowed between two buffer solutions is Δ 2 pH.
The “Configuration - Calibration” menu is extended as follows:
+ Calibration
Cal mode
Calimatic | Manual | Data entry | SOP cal
Adapt SOP cal
Cal points
1-point | 2-point | 3-point
Buffer 1
Buffer set
Mettler-Toledo
Knick CaliMat
Ciba
NIST technical
NIST standard
Hach
WTW
Hamilton
Reagecon
DIN 19267
User buffer 1
2.00/4.01/7.00/9.21
2.00/4.00/7.00/9.00/12.00
2.06/4.00/7.00/10.00
1.68/4.00/7.00/10.01/12.46
1.679/4.006/6.865/9.180
4.01/7.00/10.01/12.00
2.00/4.01/7.00/10.00
2.00/4.01/7.00/10.01/12.00
2.00/4.00/7.00/9.00/12.00
1.09/4.65/6.79/9.23/12.75
Buffer
Select a buffer from the selected set
Buffer 2
Select buffer set 2 and buffer (see buffer 1)
Buffer 3
Select buffer set 3 and buffer (see buffer 1)
Verification
Off | On
Delta pH
pH 0.05
(Enter maximum permitted deviation from ver-
ification buffer; exceeding this value generates
an error message)
Verification buffer
Select buffer set and buffer (see buffer 1)