ORP
ORP Calibration
1) In measuring mode, press the Menu softkey.
2) Select “Calibration” and confirm by pressing enter.
3) Select the desired “Calibration mode” and confirm by pressing enter.
4) The “TAG” menu allows the sensor TAG to be edited.
To do so, set “TAG” to On in the configuration menu (default setting: Off).
5) Enter the temperature-corrected setpoint of the calibration solution.
6) Immerse the sensor in the calibration solution and wait until the measured value
is stable.
7) Apply or Discard the ORP setpoint.
Note: Calibration is not possible when the device is connected via USB with the
Paraly SW 112 software.
Calibration
Note: To abort calibration, you can press meas at any time.