![Knick PORTAVO 904X PH User Manual Download Page 46](http://html.mh-extra.com/html/knick/portavo-904x-ph/portavo-904x-ph_user-manual_1980164046.webp)
46
pH/mV input
pH socket DIN 19 262 (13/4 mm)
pH measuring range
-2 … 16
Decimal places*
)
2 or 3
Input resistance
1 x 10
12
Ω
(0 … 35 °C/32 … 95 °F)
Input current
1 x 10
-12
A
(at RT, doubles every 10 K)
Measuring cycle
Approx. 1 s
Measurement error
1,2,3)
< 0.01 pH, TC < 0.001 pH/K
mV measuring range
–1300 … 1300 mV
Measuring cycle
Approx. 1 s
Measurement error
1,2,3)
< 0.1 % of measured value + 0.3 mV, TC < 0.03 mV/K
Temperature input
2 x Ø 4 mm for integrated or separate temperature probe
Measuring ranges
NTC30 temp. detector
-20 … 120 °C/-4 … 248°F
Pt1000 temp. detector
-40 … 250 °C/-40 … 482 °F
Measuring cycle
Approx. 1 s
Measurement error
1,2,3)
< 0.2 K (T
amb
= 23 °C/73.4 °F); TC < 25 ppm/K
Specifications
*) User-defined
1) At rated operating conditions
2) ± 1 digit
3) Plus sensor error
4) Ranges dependent on Memosens sensor
Memosens pH input
(also ISFET)
M8 socket, 4-pin, for Memosens laboratory cable
Display ranges
4)
pH
-2.00 … 16.00
mV
–2000 … 2000 mV
Temperature
-50 … 250 °C/-58 … 482 °F
Memosens input
ORP
M8 socket, 4-pin, for Memosens laboratory cable
Display ranges
4)
mV
–2000 … 2000 mV
Temperature
-50 … 250 °C/-58 … 482 °F
Specifications
Summary of Contents for PORTAVO 904X PH
Page 2: ......
Page 50: ...50 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Page 51: ...51 pH 02 Knick CaliMat Nominal values in bold C pH Buffer Tables ...
Page 52: ...52 pH 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated Buffer Tables ...
Page 53: ...53 pH 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added Buffer Tables ...
Page 55: ...55 pH 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH Buffer Tables ...
Page 56: ...56 pH 07 WTW Technical Buffers Nominal values in bold C pH Buffer Tables ...
Page 57: ...57 pH 08 Hamilton Nominal values in bold C pH Buffer Tables ...
Page 58: ...58 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Page 59: ...59 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...
Page 66: ...66 V Value added features 8 Viewing recorded data 32 Viewing the logger data 32 Index ...